NanoPhotometer® NP80 Tech Specs

Microvolume and Cuvette Functionality | Starting at 0.3 µl | Scan Range 200 – 900 nm

General Specifications

Dimensions: 20 cm x 20 cm x 12 cm

Weight: 3.8 – 5.2 kg depending on configuration

Processing Power/Software

  • Linux based NPOS
  • Intel Celeron dual core 2.4 GHz
  • 32 GB Internal Data Storage
  • 2x USB A, USB B
  • HDMI, Ethernet, WiFi
  • Windows 8, 10 (32 & 64 bit)
  • OS X (Intel x86 and Apple M1)
  • iOS and Android OS

Display

  • 1024 x 600 pixels
  • Glove compatible touchscreen

Built-in Battery Pack

Optional rechargeable lithium ion battery:
95 Wh, 6.6 Ah, 8 h
Min. charging cycles: 800

NanoVolume
Performance

Detection Range dsDNA 1 – 16,500 ng/µl
Detection Range BSA 0.03 – 478 mg/ml
Minimum Sample Size 0.3 – 2 µl
Photometric Range (10 mm equivalent) 0.02 – 330 A
Path Length 0.67 and 0.07 mm
Dilution Factor 15 and 140
Vortex 2,800 rpm; tube size up to 2.0 ml

Cuvette
Performance

Detection Range dsDNA 0.1 – 130 ng/µl
Detection Range BSA 0.003 – 3.7 mg/ml
Photometric Range 0 – 2.6 A
Center Height (Z-Height) 8.5 mm
Cell Types Outside dimension 12.5 x 12.5 mm
Heating 37°C ± 0.5°C

Optical
Specifications

Wavelength Scan Range 200 – 900 nm
Measure Time For Full Scan Range 2.5 – 4 seconds
Wavelength Reproducibility ± 0.2 nm
Wavelength Accuracy ± 0.75 nm
Bandwidth < 1.5 nm
Absorbance Reproducibility (Cuvette): < 0.002 A @ 0 – 0.3 A @ 280 nm CV < 1% @ 0.3 – 2.0 A @ 280 nm (Lid 15): < 0.002 A @ 0 – 0.3 A @ 280 nm CV < 1% @ 0.3 – 1.7 A @ 280 nm
Absorbance Accuracy < 1.75% @ 0.7 A @ 280 nm of the reading
Stray Light < 0.5% @ 240 nm using NaI
Optical Arrangement 1 x 4096 CMOS Array
Lamp Xenon flash lamp
Lifetime 109 flashes, up to 10 years

General
Specifications

Operating Voltage 90 – 250 V ± 10%, 50/60 Hz, 90 W, 18/19 VDC
Certification CE, IEC 61010-1:2012 and EN 61326-1:2013
Battery Certification IEC 62133 and UN38.3 transport test
Security Slot for Kensington lock
Scroll to Top